ProteoSpin™ Urine Protein Concentration Kits
For research use only and NOT intended for in vitro diagnostics.
ProteoSpin™ Urine Protein Concentration Kits
Register today to receive an exclusive 15% off* on your first order.
Supporting Data
Kit Specifications
|
|
Maximum Urine Input Volume |
1 mL
|
Maximum Recovered Protein |
200 ÎĽg
|
Elution Volume |
100 ÎĽL
|
Time to Process 12 Samples |
20 minutes
|
Storage Conditions
All solutions should be kept tightly sealed and stored at room temperature. Once opened, the solutions should be stored at 4°C. This kit is stable for 2 years after the date of shipment.
Component | Cat. 17400 (25 preps) | Cat. 52300 (10 preps) | Cat. 21600 (4 preps) |
---|---|---|---|
Wash Solution C | 60 mL | 60 mL | 130 mL |
Binding Buffer A | 4 mL | 4 mL | 8 mL |
Elution Buffer C | 8 mL | 30 mL | 30 mL |
Protein Neutralizer | 4 mL | 4 mL | 4 mL |
Micro Spin Columns | 25 | - | - |
Midi Spin Columns (assembled with collection tubes) | - | 10 | - |
Maxi Spin Columns (assembled with collection tubes) | - | - | 4 |
Collection Tubes | 25 | - | - |
Elution Tubes (1.7 mL) | 25 | - | - |
Midi Elution Tubes | - | 10 | - |
Elution Tubes (50 mL) | - | - | 4 |
Product Insert | 1 | 1 | 1 |
Documentation
FAQs
Micro, Midi, Maxi
-
Centrifugation speed was too low.
Check the centrifuge to ensure that it is capable of generating 3,300 x g (micro kit), or 1,000 x g (midi and kits). Sufficient centrifugal force is required to move the liquid phase through the resin. Centrifugation speeds may be increased to 6,700 x g (micro kit), or 2,000 x g (midi and maxi kits), but this speed should not be exceeded.
-
Inadequate spin time.
Spin additional 1 minute (micro kit), or 3 minutes (midi and maxi kits) to ensure that the liquid is able to flow completely through the column.
- Incorrect pH adjustment of urine sample.
Depending on a person’s acid-base status, the starting pH of the urine may range from 4.5 to 8. Therefore, it is important that the proper amount of Binding Buffer A be added to the urine sample in order to adjust the pH to 3.5 - 4 prior to loading onto the column. Especially with large volumes for midi and maxi kits, it is important to verify this and if necessary, add more Binding Buffer A to adjust the sample to pH 3.5 - 4.
- Initial volume of sample applied to column was too low.
Ensure that 1 mL (micro kit), or 5 mL (midi kit), or 20 mL (maxi kit) of the pH-adjusted urine sample is loaded onto the column in order to capture a large portion of the proteins present in the sample.
- Eluted protein solution was not neutralized.
Please refer to individual product manual for correct volume of Protein Neutralizer to be used in order to adjust the pH to neutral. Some proteins are sensitive to high pH, such as the Elution Buffer C at pH 12.5.
- Eluted protein solution was not neutralized quickly enough.
If eluted protein is not neutralized immediately, degradation will occur. We strongly suggest adding Protein Neutralizer to lower the pH.