2019-nCoV TaqMan RT-PCR Detection Kit

For research use only and NOT intended for in vitro diagnostics.
2019-nCoV TaqMan RT-PCR Detection Kit
Register today to receive an exclusive 15% off* on your first order.
Supporting Data
Figure 1. Analytical sensitivity for the detection of 2 SARS-CoV-2 nucleocapsid target genes and the RNase P internal Control target as recommended by the Centers for Disease Control and Prevention (CDC). Panel A represents the 2019-nCoV_N1 target, Panel B represents the 2019-nCoV_N2 target and Panel C represent the RNase P internal Control target. The analytical sensitivity of Norgen’s 2019-nCoV TaqMan RT-PCR Kit was determined by analyzing a dilution series of quantified Norgen’s 2019-nCoV RT-PCR Positive Control (Cat. PC67102) in triplicate (n=3) based on the use of TaqMan® technology. The linear range of Norgen’s 2019-nCoV TaqMan RT-PCR Kit has been determined to detect as low as 2 copies/µL with a confidence ≥ 90%.
Table 1. Operator Reproducibility. Three different operator tested 3 different concentrations in triplicate (n=3) from the 3 SARS-CoV-2 nucleocapsid target genes and the RNase P internal Control using Norgen’s 2019-nCoV TaqMan RT-PCR Kit to assess operator reproducibility. For each test the same kit Lot# and the same PCR instrument was used. The data in Table 1 showing the average Ct. values for the 3 SARS-CoV-2 nucleocapsid targets combined from the 3 different operators.
Table 2. Inter-instrument Reproducibility. Inter-instrument reproducibility was determined by running 3 different concentrations in triplicate (n=3) from the 3 SARS-CoV-2 nucleocapsid target genes and the RNase P internal Control using Norgen’s 2019-nCoV TaqMan RT-PCR Kit across 2 different qPCR instruments. For each test the same kit Lot# was used. The data in Table 2 showing the average Ct. values for the 3 SARS-CoV-2 nucleocapsid targets combined.
Table 3. Diagnostic Accuracy. Clinical evaluation for Norgen’s 2019-nCoV TaqMan RT-PCR kit was assessed by conducting a blind test with 10 different samples being either spiked or non-spiked with Norgen’s 2019-nCoV RT-PCR Positive Control (Cat. PC67102) to determine the Positive Predictive Value (PPV), Negative Predictive Value (NPV), Sensitivity, Specificity and the overall Diagnostic Accuracy.
Storage Conditions and Product Stability
All kit components should be stored at -20°C upon arrival. Repeated thawing and freezing (> 2 x) of the Master Mix, Control Master Mix and Positive Control should be avoided, as this may affect the performance of the assay. If the reagents are to be used only intermittently, they should be frozen in aliquots. All reagents can be stored for 1 year at -20°C without showing any reduction in performance.
Component | Cat. TM67100 (50 preps) | Cat. TM67120 (500 preps) |
---|---|---|
2019-nCoV_N1 Probe/Primer Mix | 80 μL | 850 μL |
2019-nCoV_N2 Probe/Primer Mix | 80 μL | 850 μL |
RNAse P Probe/Primer Mix | 80 μL | 850 μL |
2019-nCoV RT-PCR Positive Control | 50 μL | 500 μL |
2X One-Step RT-PCR Master Mix | 2 x 1 mL | 20 x 1 mL |
Nuclease-Free Water (Negative Control) | 1.25 mL | 10 x 1.25 mL |
Product Insert | 1 | 1 |